The role of thick soft lenses
In central ectasia and more severe forms, the effectiveness of visual compensation depends on the formation of a tear meniscus between the cornea and the lens. In soft lenses, this effect is achieved by specifically increasing the central thickness (CT), which in turn increases the flexural strength of the hydrophilic polymer.
The application protocol for thickened soft lenses is particularly suitable in more advanced cases of keratoconus (KC), when visual recovery is most affected by higher-order aberrations, such as vertical coma.
Construction features
Recommended polymers
To ensure minimal hypoxic stress, SP-K lenses are manufactured in silicone hydrogel (hydrogel materials 54, 59, 68% H2O available on request).
To ensure minimal hypoxic stress, SP-K lenses are manufactured in silicone hydrogel (hydrogel materials 54, 59, 68% H2O available on request).
Lens wearer selection
Intermediate ectasia, in cases of discomfort with corneal (R)GP lenses, difficult management of scleral lenses and morphology unsuitable for hybrid or twin lenses.
Intermediate ectasia, in cases of discomfort with corneal (R)GP lenses, difficult management of scleral lenses and morphology unsuitable for hybrid or twin lenses.
Refractive criteria for selection
- BCVA < 5-6/10 (visual recovery achieved with the best sphero-cylindrical ophthalmic correction).
- Inadequate visual quality, affected by residual aberrations (HOA) and the perception of ghost images, halos and double vision.
- Corneal topography: ectasia with central apex and medium-advanced progression.
Application procedures
- Select the SP-K model (1,2,3) according to the degree of KC progression in the topography; take the BC closest to the K-flat value in the SIM-K.
- 30 minutes after insertion, assess the movement and coverage ratio (TD vs HVID); if the lac is flat, modify the sagittal by adjusting the peripheral alignment skirt (step ±1).
- Check the rotation of the landmarks and perform subjective over-refraction (sphere-cylinder-axis).
- Upon removal, monitor the effects of mechanical interaction (topography) and the physiological response in laf (biomicroscopy).
Order of the Definitive Lens
Indicate any changes in parameters (BC, TD) and polymer (%H2O). Consider the over-refraction data and, in the event of residual astigmatism, order lenses with an external toric (ET) module.
Depending on the water content of the material, frequent replacement of the lenses is recommended.
Depending on the water content of the material, frequent replacement of the lenses is recommended.