High design standards
Manufactured in multiple geometries, soft toric prescription lenses offer excellent results in compensating for simple or compound astigmatism, and in combination with presbyopia.
Internal Toric
The posterior toric profile promotes stability and orientation of the lenses in the direction of the main meridians.
The base curve is calculated in relation to astigmatism, while the stabilisation system is chosen based on eyelid shape and clinical needs.
Indications for use
Corneal astigmatism SR/CR: > 0.75/1.00 D
Cylindrical refraction: significant
Stabilisation: Ballast prism, Dynamic
The base curve is calculated in relation to astigmatism, while the stabilisation system is chosen based on eyelid shape and clinical needs.
Indications for use
Corneal astigmatism SR/CR: > 0.75/1.00 D
Cylindrical refraction: significant
Stabilisation: Ballast prism, Dynamic
External Toric
The spherical inner profile is associated with frontal toric correction and the orientation is determined by the stabilisation system.
Ideal for compensating for low astigmatism according to (SR) and against the rule (CR) and also advantageous in irregular shapes.
Indications for use
Corneal astigmatism SR/CR < 0.75/1.00 D
Cylindrical refraction: low
Stabilisation: Ballast prism, Dynamic
Ideal for compensating for low astigmatism according to (SR) and against the rule (CR) and also advantageous in irregular shapes.
Indications for use
Corneal astigmatism SR/CR < 0.75/1.00 D
Cylindrical refraction: low
Stabilisation: Ballast prism, Dynamic
Bi-Toric
BT geometries combine a toric inner profile with an outer cylindrical component; the rear surface allows corneal alignment, while the front toricity compensates for residual astigmatism.
Indications for use
Corneal astigmatism: high or irregular
Cylindrical refraction: high value
Stabilisation: Ballast prism, Dynamic
Indications for use
Corneal astigmatism: high or irregular
Cylindrical refraction: high value
Stabilisation: Ballast prism, Dynamic
Construction parameters
The definition of parameters is subject to technical and clinical aspects that are entirely the responsibility of the professional. Therefore, all recommendations are to be considered guidelines and must be verified during application tests.
TSLAC stabilisations
When fitting toric lenses, there are several factors that influence success:
- eyelid aperture, orientation and tension
- blinking direction
- type, degree and axis of astigmatism.
In these delicate balances, stabilisation systems play a fundamental role in the corrective result.
Ballast prism
The asymmetrical prismatic system (approximately 1.25DP) stabilises thanks to eyelid interaction and the force of gravity.
Landmarks: 2 nasal lines.
Indications for use
Astigmatism of any degree.
High stability of correction required.
Landmarks: 2 nasal lines.
Indications for use
Astigmatism of any degree.
High stability of correction required.
TSLAC truncation (only on Internal Toric)
Made only on Prismatic Internal Torics, according to TS know-how; thanks to its alignment with the eyelid rim, it promotes stability and minimises axis rotations.
Truncation: adjustable by degrees.
Indications for use
High astigmatism: > 3.00 D
Recommended polymers: low and medium hydrophilicity <50%
Truncation: adjustable by degrees.
Indications for use
High astigmatism: > 3.00 D
Recommended polymers: low and medium hydrophilicity <50%
Dynamic Slab-Off
Optimised thickness reduction, offering thinner lenses with high comfort and no prismatic effects; ideal for symmetrical eyelid openings.
BC: the calculated theoretical radius is narrow by 0.20 mm.
Markers: 3 in OD, 2 in OS.
Indications for use
Astigmatism of any degree
Requirements: high comfort and greater Dk/t.
BC: the calculated theoretical radius is narrow by 0.20 mm.
Markers: 3 in OD, 2 in OS.
Indications for use
Astigmatism of any degree
Requirements: high comfort and greater Dk/t.